FDA establishment 3010900508, Santa Barbara, California, United States. Part of NeuroField's FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations.
Inspections: 3
Official action indicated: 2
Warning letters: 1
Inspections by fiscal year
FY2018 to FY2027 to date; 0 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
21 CFR 820.30(j). Design history file The design history file was not established.
21 CFR 820.70(i). Software validation for automated processes Software used as part of the quality system has not been validated for its intended use according to an established protocol.
21 CFR 820.50. Purchasing controls, Lack of or inadequate procedures Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
21 CFR 820.40. Procedures not adequately established or maintained Document control procedures have not been adequately established and maintained.
21 CFR 820.90(a). Nonconforming product, Lack of or inadequate procedures Procedures have not been adequately established to control product that does not conform to specified requirements.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been adequately established.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been adequately established.
21 CFR 820.20(c). Management review - Lack of or inadequate procedures Procedures for management review have not been adequately established.
21 CFR 820.22. Quality audits - Lack of or inadequate procedures Procedures for quality audits have not been adequately established.
21 CFR 820.72(a). Calibration, Inspection, etc. Procedures Lack of or Inadequ Procedures to ensure equipment is routinely calibrated, inspected, checked and maintained have not been established.
21 CFR 820.50. Purchasing controls, Lack of or inadequate procedures Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
21 CFR 820.50(a)(3). Acceptable supplier records, inadequate records Records of acceptable suppliers and contractors have not been established.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
21 CFR 820.30(g). Design validation - software validation not performed Validation of device software was not performed.
21 CFR 820.70(i). Software validation for automated processes Software used as part of production has not been validated for its intended use according to an established protocol.
21 CFR 820.20(b). Lack of or inadequate organizational structure The organizational structure has not been adequately established and maintained to ensure that devices are produced in accordance with 21 CFR 820.
21 CFR 820.90(a). Nonconforming product, Lack of or inadequate procedures Procedures have not been adequately established to control product that does not conform to specified requirements.
21 CFR 820.72(a). Calibration, Inspection, etc. Procedures Lack of or Inadequ Procedures to ensure equipment is routinely calibrated, inspected, checked and maintained have not been adequately established.
21 CFR 820.75(a). Lack of or inadequate process validation A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Voluntary Action Indicated (VAI)
Warning letters
Letters naming this establishment; each opens the letter
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm. Form 483 observations are read by Arca from the 483s FDA posts in its FOIA reading room, most of them scans: the first sentence of each observation, with redactions as FDA made them. FDA posts 483s for a minority of inspections.FDA Data Dashboard, warning letters and enforcement reports