FDA establishment 3010089756, Beaverton, Oregon, United States. Part of US Vascular's FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations.
Inspections: 6
Official action indicated: 2
Warning letters: 1
Inspections by fiscal year
FY2018 to FY2027 to date; 1 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
21 CFR 820.30(j). Design history file The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been established.
21 CFR 820.22. Quality audits - Lack of or inadequate procedures Procedures for quality audits have not been established.
21 CFR 820.90(a). Nonconforming product, Lack of or inadequate procedures Procedures have not been established to control product that does not conform to specified requirements.
21 CFR 820.184. Lack of or inadequate DHR procedures Procedures for device history records have not been established.
21 CFR 803.17. Lack of Written MDR Procedures Written MDR procedures have not been implemented.
21 CFR 820.80(a). Lack of or inadequate procedures - Acceptance activities Procedures for acceptance activities have not been established.
21 CFR 820.40. Lack of procedures, or not maintained Document control procedures have not been established.
21 CFR 820.50. Purchasing controls, Lack of or inadequate procedures Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
21 CFR 820.30(i). Design changes - Lack of or Inadequate Procedures Procedures for design change have not been established.
21 CFR 820.181. DMR - not or inadequately maintained A device master record has not been maintained.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
21 CFR 820.90(a). Nonconforming product, Lack of or inadequate procedures Procedures have not been adequately established to control product that does not conform to specified requirements.
21 CFR 820.30(j). Design history file The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.
21 CFR 803.17(a)(3). Lack of System for Timely Submission of Reports The written MDR procedure does not include an internal system which provides for timely transmission of complete medical device reports to FDA.
21 CFR 820.80(a). Lack of or inadequate procedures - Acceptance activities Procedures for acceptance activities have not been adequately established.
21 CFR 820.181. DMR - not or inadequately maintained A device master record has not been adequately maintained.
21 CFR 820.40. Procedures not adequately established or maintained Document control procedures have not been adequately established.
21 CFR 820.184. Lack of or inadequate DHR procedures Procedures for device history records have not been adequately established.
21 CFR 820.50. Purchasing controls, Lack of or inadequate procedures Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
21 CFR 820.30(i). Design changes - Lack of or Inadequate Procedures Procedures for design change have not been adequately established.
21 CFR 820.72(b). Calibration procedures - content Calibration procedures do not include specific directions and limits for accuracy and precision.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been adequately established.
21 CFR 803.17(a)(2). Lack of System for Determining MDR Events The written MDR procedure does not include an internal system which provides for a standardized review process/procedure for determining when an event meets the criteria for reporting.
21 CFR 820.22. Quality audits - Lack of or inadequate procedures Procedures for quality audits have not been adequately established.
21 CFR 820.22. Quality audits - Lack of or inadequate procedures Procedures for quality audits have not been established.
21 CFR 820.20(a). Quality policy and objectives The quality policy and quality objectives was/were not established by management with executive responsibility.
21 CFR 820.72(a). Calibration, Inspection, etc. Procedures Lack of or Inadequ Procedures to ensure equipment is routinely calibrated and maintained have not been established.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been established.
21 CFR 820.30(a). Design control - no procedures Procedures for design control have not been established.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
21 CFR 820.181. DMR - not or inadequately maintained A device master record has not been maintained.
21 CFR 820.184. Lack of or inadequate DHR procedures Procedures for device history records have not been established.
21 CFR 820.90(a). Nonconforming product, Lack of or inadequate procedures Procedures have not been established to control product that does not conform to specified requirements.
21 CFR 803.17. Lack of Written MDR Procedures Written MDR procedures have not been developed.
21 CFR 820.80(d). Lack of or inadequate final acceptance procedures Procedures for finished device acceptance have not been established.
21 CFR 820.50. Purchasing controls, Lack of or inadequate procedures Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
Voluntary Action Indicated (VAI)
Warning letters
Letters naming this establishment; each opens the letter
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm. Form 483 observations are read by Arca from the 483s FDA posts in its FOIA reading room, most of them scans: the first sentence of each observation, with redactions as FDA made them. FDA posts 483s for a minority of inspections.FDA Data Dashboard, warning letters and enforcement reports