FDA establishment 3008069988, Irvine, California, United States. Part of A-1 Engineering's FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations.
Inspections: 4
Official action indicated: 3
Warning letters: 1
Inspections by fiscal year
FY2018 to FY2027 to date; 2 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
21 CFR 820.30(i). Design changes - Lack of or Inadequate Procedures Procedures for design change have not been established.
21 CFR 820.50. Purchasing controls, Lack of or inadequate procedures Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
21 CFR 820.90(a). Nonconforming product, Lack of or inadequate procedures Procedures have not been adequately established to control product that does not conform to specified requirements.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been adequately established.
21 CFR 803.17(b)(1). Info evaluated to determine if event was reportable The written MDR procedure does not include documentation and recordkeeping requirements for all information that was evaluated to determine if an event was reportable.
21 CFR 820.20(a). Quality policy and objectives The quality policy and quality objectives was/were not established by management with executive responsibility.
21 CFR 820.198(a). Complaints Complaint files are not adequately maintained.
21 CFR 820.184. Lack of or inadequate DHR procedures Procedures for device history records have not been adequately established.
21 CFR 820.181. DMR - not or inadequately maintained A device master record has not been maintained.
21 CFR 820.184. DHR - not or inadequately maintained A device history record has not been adequately maintained.
21 CFR 820.70(g)(1). Maintenance schedule, Lack of or inadequate schedule Schedules for the adjustment, cleaning, and other maintenance of equipment have not been established.
21 CFR 820.186. QSR The quality system record has not been maintained
21 CFR 820.50(a)(1). Documented evaluation The evaluation of potential suppliers was not documented.
21 CFR 820.50(b). Approval, inadequate purchasing data Purchasing data that clearly describe or reference specified requirements for purchased or otherwise received product and services have not been approved and established.
Official Action Indicated (OAI)
Warning letters
Letters naming this establishment; each opens the letter
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm. Form 483 observations are read by Arca from the 483s FDA posts in its FOIA reading room, most of them scans: the first sentence of each observation, with redactions as FDA made them. FDA posts 483s for a minority of inspections.FDA Data Dashboard, warning letters and enforcement reports