FDA establishment 3007899424, College Station, Texas, United States. Part of CardioQuip's FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations and its recalls since 2016.
Inspections: 3
Official action indicated: 1
Warning letters: 1
Recalls since 2016: 4
Inspections by fiscal year
FY2018 to FY2027 to date; 1 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
21 CFR 820.30(d). Design output - documentation Design output was not adequately documented before release.
21 CFR 820.198(d). Records of MDR Investigation Investigation records of MDR reportable complaints do not include required information.
21 CFR 820.200(a). Servicing - Lack of or inadequate procedures Procedures or instructions for performing servicing activities and verifying that servicing meets specified requirements have not been adequately established.
21 CFR 803.17. Lack of Written MDR Procedures Written MDR procedures have not been developed and implemented.
21 CFR 820.30(c). Design input - Lack of or inadequate procedures Procedures for design input have not been adequately established.
21 CFR 820.30(i). Design changes - Lack of or Inadequate Procedures Procedures for design change have not been adequately established.
21 CFR 820.198(b). Review and evaluation for investigation Not all complaints have been adequately reviewed and evaluated to determine whether an investigation is necessary.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
21 CFR 820.100(b). Documentation Corrective and preventive action activities and/or results have not been adequately documented.
21 CFR 820.30(g). Design Validation - Risk analysis not performed/inadequate Risk analysis is inadequate and is incomplete.
21 CFR 820.30(g). Design validation- Lack of or inadequate procedures Procedures for design validation have not been adequately established.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been adequately established.
21 CFR 820.50(a)(3). Acceptable supplier records, inadequate records Records of acceptable contractors have not been adequately established.
21 CFR 820.30(i). Design changes - Lack of or Inadequate Procedures Procedures for design change have not been adequately established.
21 CFR 820.75(b). Lack/Inad procedure-Monitoring/Control of Validated Proces Procedures for monitoring and control of process parameters for a validated process have not been adequately established.
21 CFR 803.17. Lack of Written MDR Procedures Written MDR procedures have not been developed and implemented.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm. Form 483 observations are read by Arca from the 483s FDA posts in its FOIA reading room, most of them scans: the first sentence of each observation, with redactions as FDA made them. FDA posts 483s for a minority of inspections.FDA Data Dashboard, warning letters and enforcement reports