FDA establishment 3006749547, Tamarac, Florida, United States. Part of Richline Group's FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations.
Inspections: 3
Official action indicated: 2
Warning letters: 1
Inspections by fiscal year
FY2018 to FY2027 to date; 2 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
21 CFR 820.80(b). Incoming acceptance records, documentation Acceptance or rejection of incoming product was not documented.
21 CFR 820.250(b). Sampling methods - Lack of or inadequate procedures Procedures to ensure sampling methods are adequate for their intended use have not been adequately established.
21 CFR 820.181. DMR - not or inadequately maintained A device master record has not been adequately maintained.
21 CFR 820.90(a). Nonconforming product control Products that do not conform to specifications are not adequately controlled.
21 CFR 820.50(a). Evaluation of suppliers, contractors, etc., requirements Requirements that must be met by suppliers and consultants have not been adequately established.
21 CFR 820.72(a). Calibration, Inspection, etc. Procedures Lack of or Inadequ Procedures to ensure equipment is routinely calibrated, inspected, checked and maintained have not been adequately established.
21 CFR 820.184. Lack of or inadequate DHR procedures Procedures for device history records have not been adequately established.
21 CFR 820.75(a). Lack of or inadequate process validation A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
21 CFR 820.30(a). Design control - no procedures Procedures for design control have not been established.
21 CFR 820.60. Identification procedures, Lack of or inadequate procedures Procedures for identifying product during all stages of receipt, production, distribution, and installation have not been adequately established.
21 CFR 820.70(e). Contamination control, Lack of or inadequate procedures Procedures to prevent contamination of equipment or product by substances that may have an adverse effect on product quality have not been adequately established.
21 CFR 801.20(a). Label to bear a unique device identifier The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.
21 CFR 820.25(b). Training - Lack of or inadequate procedures Procedures for training and identifying training needs have not been adequately established.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been adequately established.
21 CFR 830.300(a). Devices subject to device identification GUDID data submission requirements. The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.
21 CFR 820.70(i). Software validation for automated processes Software used as part of production has not been validated for its intended use according to an established protocol.
Official Action Indicated (OAI)
Jan 12, 2024
Devices · Compliance: Devices
16 cited references
21 CFR 820.80(b). Incoming acceptance records, documentation Acceptance or rejection of incoming product was not documented.
21 CFR 820.70(g)(1). Maintenance schedule, Lack of or inadequate schedule Schedules for the adjustment, cleaning, and other maintenance of equipment have not been adequately established.
21 CFR 820.50(b). Supplier notification of changes There is no agreement with suppliers and contractors to notify you of changes in the product or service.
21 CFR 820.184. Lack of or inadequate DHR procedures Procedures for device history records have not been adequately established.
21 CFR 820.72(a). Calibration, Inspection, etc. Procedures Lack of or Inadequ Procedures to ensure equipment is routinely calibrated, inspected, checked and maintained have not been established.
21 CFR 820.75(a). Lack of or inadequate process validation A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
21 CFR 820.90(b)(2). Product rework documentation, DHR {see also 820.184} Rework and reevaluation activities have not been documented in the device history record.
21 CFR 820.25(b). Training - Lack of or inadequate procedures Procedures for training and identifying training needs have not been established.
21 CFR 820.20(b)(3). Management representative No management representative had been appointed to ensure that quality system requirements are met, and to report to management on the performance of the quality system.
21 CFR 820.22. Quality Audit/Reaudit - conducted Quality audits have not been performed.
21 CFR 820.20(c). Management review - defined interval, sufficient frequency Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system.
21 CFR 820.70(i). Software validation for automated processes Software used as part of production has not been validated for its intended use according to an established protocol. Electronic records are used, but they do not meet systems validation, system access limitation and audit trail requirements to ensure that they are trustworthy, reliable and generally equivalent to paper records.
21 CFR 820.20(a). Management ensuring quality policy is understood Management with executive responsibility has not ensured that the quality policy is understood, implemented and maintained at all levels of the organization.
21 CFR 820.70(e). Contamination control, Lack of or inadequate procedures Procedures to prevent contamination of equipment or product by substances that may have an adverse effect on product quality have not been adequately established.
21 CFR 820.181. DMR - not or inadequately maintained A device master record has not been adequately maintained.
21 CFR 820.40(a). Document review, approval documentation The documentation of approval of documents does not include the document approval date and the signature of the approving official.
Official Action Indicated (OAI)
Nov 3, 2016
Drugs · Drug Quality Assurance
6 cited references
21 CFR 211.67(a). Cleaning / Sanitizing / Maintenance Equipment and utensils are not cleaned, maintained and sanitized at appropriate intervals to prevent contamination that would alter the safety, identity, strength, quality or purity of the drug product.
21 CFR 211.113(a). Procedures for non-sterile drug products Procedures designed to prevent objectionable microorganisms in drug products not required to be sterile are not established, written and followed.
21 CFR 211.192. Investigations of discrepancies, failures There is a failure to thoroughly review the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed.
21 CFR 211.186(a). Written procedures followed Procedures for the preparation of master production and control records are not described in a written procedure and followed.
21 CFR 211.63. Equipment Design, Size and Location Equipment used in the manufacture, processing, packing or holding of drug products is not of appropriate design to facilitate operations for its intended use and cleaning and maintenance.
21 CFR 211.188. Prepared for each batch, include complete information Batch production and control records do not include complete information relating to the production and control of each batch.
Voluntary Action Indicated (VAI)
Warning letters
Letters naming this establishment; each opens the letter
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm. Form 483 observations are read by Arca from the 483s FDA posts in its FOIA reading room, most of them scans: the first sentence of each observation, with redactions as FDA made them. FDA posts 483s for a minority of inspections.FDA Data Dashboard, warning letters and enforcement reports