FDA establishment 3005643513, Nepean, Canada. Part of Abbott Point of Care Canada's FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations.
Inspections: 8
Official action indicated: 1
Warning letters: 1
Import refusals since 2014: 1
Inspections by fiscal year
FY2018 to FY2027 to date; 1 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
21 CFR 820.30(g). Design validation - production units The design was not validated under defined operating conditions and using initial production units, lots or batches or their equivalents.
21 CFR 820.20(c). Management review - Lack of or inadequate procedures Procedures for management review are not implemented.
21 CFR 820.100(a)(2). Investigation Certain indicators of nonconformities are not investigated to determine the cause of the nonconformity.
21 CFR 820.50. Purchasing controls, Lack of or inadequate procedures Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements were not implemented.
21 CFR 820.25(b). Training records Employee training is not documented.
21 CFR 820.70(c). Environmental control Lack of or inadequate procedures Appropriate procedures have not been defined for controlling environmental conditions.
21 CFR 820.22. Quality audits - Lack of or inadequate procedures Procedures for conducting quality audits were not defined.
21 CFR 820.70(a)(1). Process control instructions Process controls are not conducted in accordance with documented instructions and standard operating procedures.
Voluntary Action Indicated (VAI)
Nov 3, 2008
Devices · Postmarket Assurance: Devices
None posted
Voluntary Action Indicated (VAI)
Warning letters
Letters naming this establishment; each opens the letter
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm. Form 483 observations are read by Arca from the 483s FDA posts in its FOIA reading room, most of them scans: the first sentence of each observation, with redactions as FDA made them. FDA posts 483s for a minority of inspections.FDA Data Dashboard, warning letters and enforcement reports