FDA establishment 3005612924, Lindsay, Canada. Part of Cardiomed Supplies's FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations, and the shipments FDA refused at the border.
21 CFR 820.198(e). Records of complaint investigation Records of complaint investigations do not include required information.
21 CFR 820.198(d). Evaluation, timeliness, identification Complaints representing events that are MDR reportable were not promptly reviewed, evaluated, and investigated by a designated individual and clearly identified.
21 CFR 820.22. Quality Audit/Reaudit - conducted Quality audits have not been performed.
21 CFR 820.100(b). Documentation Corrective and preventive action activities and/or results have not been adequately documented.
Voluntary Action Indicated (VAI)
Warning letters
Letters naming this establishment; each opens the letter
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm. Form 483 observations are read by Arca from the 483s FDA posts in its FOIA reading room, most of them scans: the first sentence of each observation, with redactions as FDA made them. FDA posts 483s for a minority of inspections.FDA Data Dashboard, warning letters and enforcement reports