FDA establishment 3004068499, Itasca, Illinois, United States. Part of Organ Recovery Systems's FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations and its recalls since 2016.
Inspections: 5
Official action indicated: 1
Warning letters: 1
Recalls since 2016: 2
Inspections by fiscal year
FY2018 to FY2027 to date; 0 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
21 CFR 820.75(a). Lack of or inadequate process validation A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
21 CFR 820.160(a). Lack of or inadequate procedures for distribution Procedures for control and distribution of finished devices have not been established.
21 CFR 820.50. Purchasing controls, Lack of or inadequate procedures Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
21 CFR 820.30(g). Design validation - documentation The results of design validation, including identification of the design, method(s), the date and the individual(s) performing validation, were not adequately documented in the design history file.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been adequately established.
21 CFR 820.50(a)(3). Acceptable supplier records, inadequate records Records of acceptable suppliers have not been adequately established.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been adequately established.
21 CFR 820.100(b). Documentation Corrective and preventive action activities and/or results have not been adequately documented.
21 CFR 803.17(a)(1). Lack of System for Event Evaluations The written MDR Procedure does not include an internal system which provides for the timely and effective identification of events that may be subject to medical device reporting requirements.
21 CFR 820.22. Quality audits - Lack of or inadequate procedures Procedures for quality audits have not been adequately established.
No Action Indicated (NAI)
Nov 8, 2011
Devices · Compliance: Devices
None posted
Voluntary Action Indicated (VAI)
Dec 9, 2009
Devices · Compliance: Devices
1 cited reference
21 CFR 820.30(i). Design changes - Lack of or Inadequate Procedures Procedures were not followed for the documentation, validation or verification, review, and approval of design changes before their implementation.
Voluntary Action Indicated (VAI)
Warning letters
Letters naming this establishment; each opens the letter
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm. Form 483 observations are read by Arca from the 483s FDA posts in its FOIA reading room, most of them scans: the first sentence of each observation, with redactions as FDA made them. FDA posts 483s for a minority of inspections.FDA Data Dashboard, warning letters and enforcement reports