FDA establishment 3002900199, Brisbane, California, United States. Part of Cutera's FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations.
21 CFR 820.70(i). Software validation for automated processes Software used as part of the quality system has not been adequately validated for its intended use according to an established protocol.
Voluntary Action Indicated (VAI)
Nov 5, 2012
Devices · Postmarket Assurance: DevicesDevices · Compliance: Devices; Devices · Radiation Control and Health Safety Act
None posted
No Action Indicated (NAI)
Aug 30, 2011
Devices · Postmarket Assurance: DevicesDevices · Radiation Control and Health Safety Act; Devices · Compliance: Devices
5 cited references
21 CFR 806.10(b). Time to report - 10 days A report of the required information regarding device correction and removal actions was not sent to FDA within 10 days of initiating the correction or removal.
21 CFR 803.11. Failure to use FDA Form 3500A An MDR adverse event report was submitted on a form other than FDA Form 3500A (MEDWATCH form) or an approved electronic equivalent.
21 CFR 820.25(b). Training - Lack of or inadequate procedures Procedures for training and identifying training needs have not been established.
21 CFR 820.198(d). Evaluation, timeliness, identification Complaints representing events that are MDR reportable were not promptly reviewed, evaluated, and investigated by a designated individual.
21 CFR 803.50(a)(1). Report of Death or Serious Injury An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Voluntary Action Indicated (VAI)
Warning letters
Letters naming this establishment; each opens the letter
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm. Form 483 observations are read by Arca from the 483s FDA posts in its FOIA reading room, most of them scans: the first sentence of each observation, with redactions as FDA made them. FDA posts 483s for a minority of inspections.FDA Data Dashboard, warning letters and enforcement reports