FDA establishment 3002807798, Hamburg, Germany. Part of Olympus Winter & Ibe's FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations, and the shipments FDA refused at the border.
Inspections: 5
Official action indicated: 1
Warning letters: 1
Import refusals since 2014: 2
Inspections by fiscal year
FY2018 to FY2027 to date; 1 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been adequately established.
21 CFR 820.50. Purchasing controls, Lack of or inadequate procedures Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
21 CFR 820.30(c). Design input - Lack of or inadequate procedures Procedures for design input have not been adequately established.
21 CFR 820.25(b). Training - Lack of or inadequate procedures Procedures for training and identifying training needs have not been established.
21 CFR 803.50(a)(2). Individual Report of Malfunction An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
21 CFR 820.30(g). Design validation - user needs and intended uses Design validation did not ensure the device conforms to defined user needs and intended uses.
Official Action Indicated (OAI)
Sep 1, 2022
Devices · Compliance: Devices
None posted
No Action Indicated (NAI)
Jan 26, 2017
Devices · Compliance: Devices
3 cited references
21 CFR 803.50(b)(1). Reporting Information Reasonably Known An MDR report submitted to FDA did not include all information that was reasonably known to the manufacturer.
21 CFR 820.100(b). Documentation Corrective and preventive action activities and/or results have not been documented.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Voluntary Action Indicated (VAI)
Dec 4, 2014
Devices · Compliance: Devices
1 cited reference
21 CFR 820.86. Acceptance status The acceptance status of product was not identified to indicate conformance or nonconformance with acceptance criteria and maintained.
Voluntary Action Indicated (VAI)
Dec 3, 2009
Devices · Compliance: Devices
None posted
Voluntary Action Indicated (VAI)
Warning letters
Letters naming this establishment; each opens the letter
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm. Form 483 observations are read by Arca from the 483s FDA posts in its FOIA reading room, most of them scans: the first sentence of each observation, with redactions as FDA made them. FDA posts 483s for a minority of inspections.FDA Data Dashboard, warning letters and enforcement reports