FDA establishment 3002807393, Numbrecht, Germany. Part of Kabe Labortechnik's FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations.
Inspections: 1
Official action indicated: 1
Warning letters: 1
Inspections by fiscal year
FY2018 to FY2027 to date; 1 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
21 CFR 820.30(a). Design control - no procedures Procedures for design control have not been established.
21 CFR 820.50. Purchasing controls, Lack of or inadequate procedures Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
21 CFR 820.75(a). Lack of or inadequate process validation A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been adequately established.
Official Action Indicated (OAI)
Warning letters
Letters naming this establishment; each opens the letter
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm. Form 483 observations are read by Arca from the 483s FDA posts in its FOIA reading room, most of them scans: the first sentence of each observation, with redactions as FDA made them. FDA posts 483s for a minority of inspections.FDA Data Dashboard, warning letters and enforcement reports