FDA establishment 3002806673, Khlong Luang, Thailand. Part of Okamoto Rubber Products's FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations.
21 CFR 806.10(a)(1). Report of risk to health A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
21 CFR 820.184. DHR content The device history record does not demonstrate that the device was manufactured in accordance with 21 CFR 820.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been adequately established.
21 CFR 820.70(e). Contamination control, Lack of or inadequate procedures Procedures to prevent contamination of equipment or product by substances that may have an adverse effect on product quality have not been established.
21 CFR 820.70(g)(1). Maintenance schedule, Lack of or inadequate schedule Schedules for the adjustment, cleaning, and other maintenance of equipment have not been established.
21 CFR 820.198(a). Complaints Complaint files are not maintained.
Official Action Indicated (OAI)
Warning letters
Letters naming this establishment; each opens the letter
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm. Form 483 observations are read by Arca from the 483s FDA posts in its FOIA reading room, most of them scans: the first sentence of each observation, with redactions as FDA made them. FDA posts 483s for a minority of inspections.FDA Data Dashboard, warning letters and enforcement reports