FDA establishment 3002553648, Berlin, Germany. Part of GLE Gesellschaft für Lichttechnische Erzeugnisse's FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations.
Inspections: 3
Official action indicated: 1
Warning letters: 1
Inspections by fiscal year
FY2018 to FY2027 to date; 0 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
21 CFR 820.90(a). Nonconforming product, Lack of or inadequate procedures Procedures have not been adequately established to control product that does not conform to specified requirements.
Voluntary Action Indicated (VAI)
Sep 28, 2017
Devices · Compliance: Devices
4 cited references
21 CFR 820.90(a). Nonconforming product, Lack of or inadequate procedures Procedures have not been adequately established to control product that does not conform to specified requirements.
21 CFR 820.250(b). Sampling plans Sampling plans are not based on valid statistical rationale.
21 CFR 820.80(e). Documentation Acceptance activities were not adequately documented.
21 CFR 820.40(a). Not approved or obsolete document retrieval Documents that were obsolete were observed at a location where they are being used.
Voluntary Action Indicated (VAI)
Jul 29, 2016
Devices · Compliance: DevicesDevices · Radiation Control and Health Safety Act
10 cited references
21 CFR 820.30(a). Design control - no procedures Procedures for design control have not been established.
21 CFR 820.70(f). Buildings Buildings do not contain sufficient space to perform necessary operations, prevent mix-ups and assure orderly handling of product.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been adequately established.
21 CFR 820.90(a). Nonconforming product control Products that do not conform to specifications are not adequately controlled.
21 CFR 820.181. DMR - not or inadequately maintained A device master record has not been maintained.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
21 CFR 820.72(b). Remedial action When test/measurement equipment was found to not meet accuracy and precision limits, inadequate action was taken to evaluate whether there was any adverse effect on the device's quality.
21 CFR 820.40(b). Document change records, maintained. Records of changes to documents were not adequately maintained.
21 CFR 820.86. Acceptance status The acceptance status of product was not identified to indicate conformance or nonconformance with acceptance criteria.
21 CFR 820.22. Quality Audits - defined intervals Quality audits were not performed to determine whether the quality system activities and results comply with quality system procedures.
Official Action Indicated (OAI)
Warning letters
Letters naming this establishment; each opens the letter
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm. Form 483 observations are read by Arca from the 483s FDA posts in its FOIA reading room, most of them scans: the first sentence of each observation, with redactions as FDA made them. FDA posts 483s for a minority of inspections.FDA Data Dashboard, warning letters and enforcement reports