FDA establishment 2242483, Clifton, New Jersey, United States. Part of Physitemp Instruments's FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations and its recalls since 2016.
Inspections: 5
Official action indicated: 1
Warning letters: 1
Recalls since 2016: 1
Inspections by fiscal year
FY2018 to FY2027 to date; 1 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
21 CFR 820.80(d). Lack of or inadequate final acceptance procedures Procedures for finished device acceptance have not been adequately established.
21 CFR 820.75(b). Lack/Inad procedure-Monitoring/Control of Validated Proces Procedures for monitoring and control of process parameters for a validated process have not been adequately established.
21 CFR 801.20(a). Label to bear a unique device identifier The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.
21 CFR 820.72(a). Equipment suitability & capability Certain test equipment is not capable of producing valid results.
21 CFR 803.17. Lack of Written MDR Procedures Written MDR procedures have not been developed, maintained and implemented.
21 CFR 820.90(a). Nonconforming product, Lack of or inadequate procedures Procedures have not been adequately established to control product that does not conform to specified requirements.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been established.
21 CFR 830.300(a). Devices subject to device identification GUDID data submission requirements. The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.
21 CFR 820.75(a). Documentation Process validation activities have not been adequately documented.
21 CFR 820.50. Purchasing controls, Lack of or inadequate procedures Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm. Form 483 observations are read by Arca from the 483s FDA posts in its FOIA reading room, most of them scans: the first sentence of each observation, with redactions as FDA made them. FDA posts 483s for a minority of inspections.FDA Data Dashboard, warning letters and enforcement reports