FDA establishment 1036223, Palm City, Florida, United States. Part of Awareness Technology's FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations.
Inspections: 4
Official action indicated: 0
Warning letters: 0
Inspections by fiscal year
FY2018 to FY2027 to date; 0 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
21 CFR 820.70(g)(1). Maintenance schedule, Lack of or inadequate schedule Schedules for the adjustment, cleaning, and other maintenance of equipment have not been adequately established.
21 CFR 820.198(e). Records of complaint investigation Records of complaint investigations do not include required information.
21 CFR 820.100(b). Documentation Corrective and preventive action activities and/or results have not been adequately documented.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
21 CFR 820.30(g). Design validation- Lack of or inadequate procedures Procedures for design validation have not been adequately established.
21 CFR 820.75(a). Lack of or inadequate process validation A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
21 CFR 820.70(i). Software validation for automated processes Software used as part of production has not been validated for its intended use according to an established protocol.
21 CFR 820.80(e). Documentation Acceptance activities were not adequately documented.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been adequately established.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been adequately established.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
21 CFR 820.75(a). Lack of or inadequate process validation A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
21 CFR 820.70(c). Environmental control Lack of or inadequate procedures Procedures to control environmental conditions have not been adequately established.
21 CFR 803.17. Lack of Written MDR Procedures Written MDR procedures have not been maintained.
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm. Form 483 observations are read by Arca from the 483s FDA posts in its FOIA reading room, most of them scans: the first sentence of each observation, with redactions as FDA made them. FDA posts 483s for a minority of inspections.FDA Data Dashboard, warning letters and enforcement reports