Robotically-Assisted Surgical Devices - Premarket Submissions: Draft Guidance for Industry and Food and Drug Administration Staff
US Food and Drug Administration CDRH, draft guidance. In the week of Sep 21 to 27, 2026.
- Issued: Sep 25, 2026
- Comments closed: Nov 24, 2026
What it says
The draft guidance provides recommendations for premarket submissions for certain robotically-assisted surgical devices (RASDs).
This draft guidance provides recommendations for premarket submissions for certain robotically-assisted surgical devices (RASDs).
For RASD reliability testing, FDA recommends sufficient sample size for grip and cutting reliability testing to demonstrate 95% reliability with 95% confidence.
FDA recommends sufficient sample size for grip and cutting reliability testing to demonstrate 95% reliability with 95% confidence.
FDA recommends testing a minimum of three RASDs for accuracy, repeatability, and resolution testing.
FDA recommends testing a minimum of three RASDs.
FDA generally thinks device software functions for a RASD warrant an "Enhanced" Documentation Level.
We think generally the device software functions for a RASD warrant an “Enhanced” Documentation Level.
Who should care
Comments on the draft guidance are due within 60 days of publication of the Federal Register notice announcing its availability.
You should submit comments and suggestions regarding this draft document within 60 days of publication in the Federal Register of the notice announcing the availability of the draft guidance.
The guidance is relevant to training programs that cover not only surgeons but also other OR staff such as bedside assistants and surgical scrub nurses.
The training program and competency expectations should not only focus on surgeons, but also include other OR staff, such as bedside assistants and surgical scrub nurses, who interact with RASDs.
- Agency
- US Food and Drug Administration CDRH
- Type
- Guidance
- Status
- Draft
- Issued
- Sep 25, 2026
- Docket
- FDA-2026-N-9505
Status, dates, docket and comment deadline are read from FDA's own records and the docket's notice on Regulations.gov or in the Federal Register. The summary is written by a language model and checked by matching each quoted passage against the document; it is a reading aid, not the document.