Pyrogen and Endotoxins Testing: Questions and Answers: Guidance for Industry

US Food and Drug Administration CDER, CBER, CDRH, CVM, OII, final guidance. In the week of Mar 16 to 22, 2026.

  • Issued: Mar 18, 2026

Summary

Each point with the passage it rests on

What it says

  1. Firms should establish procedures for storing and handling samples for bacterial endotoxins analysis using laboratory data that demonstrate the stability of assayable endotoxins content.

    Firms should establish procedures for storing and handling (which includes product mixing) samples for bacterial endotoxins analysis using laboratory data that demonstrate the stability of assayable endotoxins content.
  2. For medical devices, the endotoxin limit is 0.5 EU/mL or 20 EU/device for products that directly or indirectly contact the cardiovascular system and lymphatic system.

    For medical devices, using the extraction volume recommendations described below, the endotoxin limit is 0.5 EU/mL or 20 EU/device for products that directly or indirectly contact the cardiovascular system and lymphatic system.
  3. For veterinary products labeled for multiple species, the endotoxins limit should be based on the maximum product dose used on the smallest species.

    For a veterinary product labeled for use in multiple species, the limit should be based on the maximum product dose used on the smallest species.

What changed

  1. The FDA has withdrawn the 1987 Guidance because it no longer reflects the Agency's current thinking on the topic.

    The Agency has withdrawn the 1987 Guidance because it no longer reflects the Agency’s current thinking on the topic.

Who should care

  1. The guidance is directed at biological product, drug, and device firms.

    This guidance provides recommendations for biological product, drug, and device firms on FDA’s current thinking concerning the testing recommendations and acceptance criteria

Sources

From FDA

Record

Agency
US Food and Drug Administration CDER, CBER, CDRH, CVM, OII
Type
Guidance
Status
Final
Issued
Mar 18, 2026

Status, dates, docket and comment deadline are read from FDA's own records and the docket's notice on Regulations.gov or in the Federal Register. The summary is written by a language model and checked by matching each quoted passage against the document; it is a reading aid, not the document.