Patient-Matched Guides for Orthopedic Implants: Guidance for Industry and Food and Drug Administration Staff
US Food and Drug Administration CDRH, final guidance. In the week of May 4 to 10, 2026.
- Issued: May 7, 2026
What it says
The guidance gives FDA's current thinking on what information should be included in regulatory submissions for patient-matched guides to orthopedic implants.
This guidance document provides recommendations regarding information that should be included in regulatory submissions for patient-matched guides to orthopedic implants.
The guidance recommends a sterility assurance level of 10^-6 for devices labeled as sterile unless the device is intended only for contact with intact skin.
You should state the sterility assurance level (SAL) of $10^{-6}$ for devices labeled as sterile unless the device is intended only for contact with intact skin.
The guidance states that clinical performance testing is generally not necessary to support regulatory evaluation of orthopedic patient-matched guides.
Clinical performance testing is generally not necessary to support regulatory evaluation of orthopedic patient-matched guides.
- Agency
- US Food and Drug Administration CDRH
- Type
- Guidance
- Status
- Final
- Issued
- May 7, 2026
- Docket
- FDA-2023-D-2370
Status, dates, docket and comment deadline are read from FDA's own records and the docket's notice on Regulations.gov or in the Federal Register. The summary is written by a language model and checked by matching each quoted passage against the document; it is a reading aid, not the document.