Paclitaxel drug-coated balloons and drug-eluting stents
UK Medicines and Healthcare products Regulatory Agency, publication. In the week of Mar 2 to 8, 2026.
- Last published: Mar 3, 2026
What it says
The Expert Advisory Group recommends withholding paclitaxel coated balloons and paclitaxel eluting stents from routine clinical use in patients with intermittent claudication until further Level 1 evidence on their safety profile is available.
The use of paclitaxel coated balloons and paclitaxel eluting stents should be withheld from routine use in patients with intermittent claudication until further Level 1 evidence on their safety profile is available.
The devices can be considered in patients with critical limb ischaemia if the benefits are felt to outweigh the risks, taking NICE Guidance into account, with enhanced patient follow-up that includes follow-up until death and reporting of serious adverse events and cause of death.
If paclitaxel coated/eluting devices are to be used there should be enhanced patient follow-up which may include telephone consultation or review in the community, where appropriate. This should include at a minimum follow up until death and reporting of serious adverse events and cause of death.
Informed consent should include a risk-benefit discussion covering the uncertainty in long-term outcomes and the current evidence of increased mortality rate.
The process of informed consent on the use of paclitaxel coated balloons and paclitaxel eluting stents should include a risk-benefit discussion with patients regarding the uncertainty in long-term outcomes with these devices, and the current evidence which indicates an increased mortality rate.
What changed
The guidance was updated in February 2024 after the MHRA reviewed more recent and comprehensive published mortality data, which is referred to as DSI/2024/001.
In February 2024, following additional review by the MHRA of more recent and comprehensive published mortality data, comparing the use of paclitaxel coated devices (PCD) to non-paclitaxel coated devices (nPCD) in the treatment of peripheral arterial disease (PAD), we have updated our guidance - see DSI/2024/001 .
Who should care
Clinicians treating peripheral arterial disease patients with these devices, and healthcare professionals who should report suspected adverse incidents, should read this guidance.
Healthcare professionals should report any suspected or actual adverse incidents involving these devices through your healthcare institution’s local incident reporting system
- Paclitaxel drug-coated balloons and drug-eluting stentsMHRA, the document itself
- Agency
- UK Medicines and Healthcare products Regulatory Agency
- Type
- Publication
- Last published
- Mar 3, 2026
Status, dates, docket and comment deadline are read from MHRA's own records. The summary is written by a language model and checked by matching each quoted passage against the document; it is a reading aid, not the document.