National Patient Safety Alert Supporting Information. Patient hoists and slings: incidents and case examples.

UK Medicines and Healthcare products Regulatory Agency, publication. In the week of Sep 14 to 20, 2026.

  • Last published: Sep 16, 2026

Summary

Each point with the passage it rests on

What it says

  1. From 1 January 2015 to 31 December 2025, the MHRA received 22 reports of incidents with a fatal outcome involving patient hoists during routine transfers.

    From 1 January 2015 to 31 December 2025, the MHRA received 22 reports of incidents with a fatal outcome involving patient hoists during routine transfers.
  2. Pre-use checks must include a "pause-and-check" step after initial load take-up, before the patient is fully lifted.

    Pre‑use checks should be capable of detecting these conditions and must include a “pause‑and‑check” step after initial load take-up, before the patient is fully lifted (for example, after slack has been removed and the sling attachments are under tension).
  3. Any hoist or sling that has been serviced or repaired must be functionally checked before being returned to service.

    Any hoist or sling that has been serviced or repaired must be functionally checked before being returned to service.
  4. Healthcare organisations and individuals should report all incidents involving patient hoists and slings to the MHRA through the Yellow Card scheme.

    Healthcare organisations and individuals should report all incidents involving patient hoists and slings to the MHRA through the Yellow Card scheme.
  5. Compatibility of hoist and sling combinations must be confirmed using the manufacturer's instructions for use and should not be assumed because a sling physically fits.

    Compatibility must be confirmed using the manufacturer’s instructions for use and should not be assumed because a sling physically fits.

Who should care

  1. The guidance is relevant to healthcare organisations and individuals involved in patient hoist and sling use, and it notes the issues may affect any patient hoist or sling.

    The issues described may affect any patient hoist or sling.

Sources

From MHRA

Record

Agency
UK Medicines and Healthcare products Regulatory Agency
Type
Publication
Last published
Sep 16, 2026