Medical Gases — Current Good Manufacturing Practice: Draft Guidance for Industry
US Food and Drug Administration CDER, CVM, OII, draft guidance. In the week of Nov 24 to 30, 2025.
- Issued: Nov 28, 2025
- Comments closed: Jan 31, 2026
What it says
Manufacturers should perform a heat-of-compression check on high-pressure cylinders during or immediately after filling, and investigate any cool or cold cylinders.
During or immediately after filling a high-pressure cylinder, manufacturers should perform a heat-of-compression check by lightly touching the exterior of the cylinder or an alternative method that verifies temperature change.
Manufacturers should reject any high-pressure cylinder whose DOT requalification date has been exceeded only after quarantining it until DOT requirements are satisfied or it is removed from inventory.
If the DOT requalification date has been exceeded, the manufacturer should quarantine the cylinder until either DOT requirements have been satisfied, or the cylinder is removed from inventory.
Who should care
The guidance is intended for manufacturers of medical gases, including those that fill, refill, or distribute them.
This guidance is intended to assist manufacturers 2 of medical gases in complying with current good manufacturing practice (CGMP) requirements.