Hemodialysis Blood Tubing Sets - Premarket Notification (510(k)) Submissions: Guidance for Industry and Food and Drug Administration Staff
US Food and Drug Administration CDRH, final guidance. In the week of Jul 20 to 26, 2026.
- Issued: Jul 23, 2026
What it says
FDA recommends that manufacturers test three or more hemodialysis tubing sets of each model for pressure leak, pump segment endurance, clamp performance, and kinking resistance, among other tests.
We recommend you perform the following tests on three or more hemodialysis tubing sets of each model you intended to market:
FDA recommends that the device be sterile with a sterility assurance level of 1 x 10-6.
We recommend the device be sterile with a sterility assurance level (SAL) of 1 x 10-6.
FDA recommends a warning that significant hemolysis of red blood cells can occur in kinked blood tubing, especially in the post-pump, arterial tubing segment.
We recommend including a warning that significant hemolysis of red blood cells can occur in kinked blood tubing, especially in the post -pump, arterial tubing segment.
FDA recommends color coding so that the arterial blood tubing connectors are red and the venous blood tubing connectors are blue.
We also recommend you color code the device so that the patient and dialyzer connectors of the arterial blood tubing are red and the patient and dialyzer connectors of the venous blood tubing are blue.
Who should care
The guidance is for industry preparing 510(k) submissions for hemodialysis blood tubing sets, and for FDA staff.
This guidance document is intended to assist industry in preparing a premarket notification submission (510(k)) for hemodialysis blood tubing set for use as the extracorporeal blood circuit during hemodialysis treatment.