Health Institution Exemption for general medical devices
UK Medicines and Healthcare products Regulatory Agency, publication. In the week of Jan 19 to 25, 2026.
- Last published: Jan 20, 2026
What it says
The MDR 2002 do not currently include an exemption for general medical devices, but health institutions should follow this guidance and consider the HIE applicable to all medical devices and accessories provided the conditions for use are met and leading practice guidelines are followed.
The MDR 2002 do not currently include an exemption for general medical devices (including software or implantable medical devices).
The HIE does not apply to devices that are manufactured within a health institution and then used outside that institution's premises or its immediate vicinity, such as devices taken home by patients.
Under the MDR 2002, devices that are manufactured within a health institution and then used outside that health institution’s premises or its immediate vicinity are not eligible for the HIE.
Before developing a device under the HIE, health institutions should assess whether commercial alternatives are available, and the HIE should only be applied where no equivalent device meets the target patient group's needs or available alternatives do not meet the required performance level.
The MHRA advises that the HIE should only be applied when: No equivalent device on the market meets the specific needs of the target patient group, or Available alternatives do not meet the required performance level.
What changed
The MHRA no longer recommends using the terms 'in-house manufacturing' or 'in-house' because of ambiguity.
In previous guidance, the term ‘in-house manufacturing’ or ‘in-house’ has been used, but the MHRA does not recommend using this moving forward to avoid ambiguity.
Who should care
This guidance is for health institutions in Great Britain (England, Wales and Scotland) that manufacture or use general medical devices under the Health Institution Exemption; Northern Ireland follows MDCG guidance and IVD devices follow separate IVD guidance.
This guidance applies to health institutions in Great Britain (England, Wales, and Scotland).
- Health Institution Exemption for general medical devicesMHRA, the document itself
- Agency
- UK Medicines and Healthcare products Regulatory Agency