Dental Composite Resin Devices - Premarket Notification (510(k)) Submissions: Guidance for Industry and Food and Drug Administration Staff

US Food and Drug Administration CDRH, final guidance. In the week of Aug 31 to Sep 6, 2026.

  • Issued: Sep 2, 2026

Summary

Each point with the passage it rests on

What it says

  1. FDA recommends non-clinical performance testing for dental composite resin devices, because descriptive characteristics alone are insufficient to support a substantial equivalence determination.

    Non-clinical performance testing is recommended for dental composite resin devices because descriptive characteristics alone are insufficient to support a substantial equivalence determination.
  2. Clinical evidence is generally unnecessary for most dental composite resin devices, but FDA may request it, for example for any statements about device performance such as longevity, tooth remineralization, or reduced decay.

    Clinical evidence is generally unnecessary for most dental composite resin devices; however, such testing may be requested in situations such as the following:
  3. FDA recommends that shelf life studies include real-time aged samples to confirm the results of accelerated aging, since dental composite resins are polymeric materials.

    Since dental composite resins are composed of polymeric materials, you should conduct testing on real-time aged samples to confirm the results of the accelerated aging study.

What changed

  1. The guidance supersedes the October 2005 final guidance on dental composite resin 510(k) submissions.

    This document supersedes Dental Composite Resin Devices – Premarket Notification [510(k)] Submissions issued October 26, 2005.

Sources

From FDA and the Federal Register

Record

Agency
US Food and Drug Administration CDRH
Type
Guidance
Status
Final
Issued
Sep 2, 2026