Arca
Skip to content
Tools
Tools

Custom-made medical devices in Great Britain

UK Medicines and Healthcare products Regulatory Agency, publication. In the week of Jul 20 to 26, 2026.

  • Last published: Jul 20, 2026

Summary

Each point with the passage it rests on

What it says

  1. A custom-made device is one manufactured specifically in accordance with a written prescription of a registered medical practitioner or other authorised person, intended for the sole use of a particular patient.

    manufactured specifically in accordance with a written prescription of a registered medical practitioner, or other person authorised to write such a prescription by virtue of their professional qualification, which gives under their responsibility, specific characteristics as to its design intended for the sole use of a particular patient
  2. Custom-made devices do not need to be UKCA marked, but they must still meet the relevant provisions of the UK MDR 2002, and manufacturers do not need a UK approved body.

    Custom-made devices and custom-made active implantable medical devices do not need to be UKCA marked when they are first placed on the market and/or put into service. However, they must meet the relevant provisions of the UK MDR 2002 that apply to them.
  3. Manufacturers must provide a statement with the device, including the patient's name or an identifier, the prescriber's name, and a statement of conformity, where the device is class IIa, class IIb, class III or an active implantable device.

    If the custom-made device in question would have been classified as either a class IIa, class II(b) or class III device, in the definition of a medical device or is an active implantable medical device then that device must be accompanied by the statement referred to
  4. Manufacturers must keep a post-market surveillance system for each device and use its data to update instructions, labelling, risk analysis and technical documentation.

    Specifically, manufacturers of custom-made devices must maintain a post-market surveillance (PMS) system for each device that is placed on the market or put into service.
  5. Manufacturers or their UK responsible person must register with the MHRA if they place custom-made devices on the market.

    A manufacturer of custom-made devices or their UK responsible person must register with the MHRA if they place the devices on the market.

Who should care

  1. Manufacturers of custom-made medical devices and active implantable devices in Great Britain should read this guidance, as should prescribers and professional users such as dentists, ocularists and orthotists who supply such devices.

    This guidance applies to medical devices placed on the market in Great Britain (England, Wales and Scotland).

Sources

From MHRA

Record

Agency
UK Medicines and Healthcare products Regulatory Agency
Type
Publication
Last published
Jul 20, 2026

Status, dates, docket and comment deadline are read from MHRA's own records. The summary is written by a language model and checked by matching each quoted passage against the document; it is a reading aid, not the document.

Esc