Cuffless Non-invasive Blood Pressure Measuring Devices – Clinical Performance Testing and Evaluation: Draft Guidance for Industry and Food and Drug Administration Staff
US Food and Drug Administration CDRH, draft guidance. In the week of Jan 19 to 25, 2026.
- Issued: Jan 23, 2026
- Comments closed: Mar 24, 2026
What it says
FDA's draft guidance provides recommendations for clinical performance testing and evaluation to support premarket submissions for cuffless non-invasive blood pressure measuring devices.
This draft guidance provides recommendations for clinical performance testing and evaluation to support premarket submissions for cuffless non-invasive blood pressure measuring devices.
FDA recommends at least 85 participants aged 21 to 50 years for devices intended for adults in that age range, including at least 30% females and at least 30% males.
We recommend that at least 85 participants between the ages 21 years and 50 years old be tested if the device is intended for use in adults between the ages 21 years and 50 years old. 5 This should include at least 30% females and 30% males.
FDA recommends that the study enroll at least 25% of participants in each Monk Skin Tone scale grouping 1-4, 5-7, and 8-10 (or equivalent scale).
For example, we recommend that the study enrolls at least 25% in each Monk Skin Tone scale 1-4, 5-7, and 8-10 (or equivalent scale).
For accuracy, FDA recommends a mean error no greater than ±5.0 mmHg and a standard deviation less than 8.0 mmHg, with the Change test evaluated only on subjects who achieve the necessary blood pressure change.
Criterion #1: A mean error less than or equal ±5.0 mmHg and a standard deviation less than 8.0 mmHg.
FDA recommends early engagement through the Q-Submission Program, and says it recommends using that program for all new devices of this type.
In general, given the novelty and variety of this device technology, FDA recommends utilizing the Q-Submission Program to obtain feedback for all new devices of this type.
Who should care
The guidance is relevant to manufacturers preparing premarket submissions for cuffless non-invasive blood pressure measuring devices under product code DXN, while cuff-based devices and neonatal ICU monitors under product code QYF are excluded.
Cuff-based devices under product code DXN and Neonatal ICU Continuous Non-Invasive Blood Pressure Monitors under product code QYF are outside the scope of this draft guidance.
- Agency
- US Food and Drug Administration CDRH
- Type
- Guidance
- Status
- Draft
- Issued
- Jan 23, 2026
- Docket
- FDA-2025-D-0851
Status, dates, docket and comment deadline are read from FDA's own records and the docket's notice on Regulations.gov or in the Federal Register. The summary is written by a language model and checked by matching each quoted passage against the document; it is a reading aid, not the document.