Clinical trials that include an in vitro diagnostic device
UK Medicines and Healthcare products Regulatory Agency, publication. In the week of Mar 9 to 15, 2026.
- Last published: Mar 9, 2026
What it says
The guidance covers in vitro diagnostic devices, including companion diagnostics, used in clinical trials of medicinal products (CTIMPs) with a trial site in Great Britain.
This guidance covers the use of in vitro diagnostic ( IVD ) devices, including companion diagnostics, in clinical trials of medicinal products (CTIMPs) with a trial site in Great Britain.
IVD devices used in GB trials must bear a UKCA or CE mark for use as intended by the manufacturer, unless a health institution exempted device applies.
IVD devices including companion diagnostic devices must bear a UKCA or CE mark for use as intended by the manufacturer in a clinical trial.
Applicants must state in the covering letter that the trial includes an IVD device and declare the device's CE or UKCA mark status or health institution exemption.
First, the applicant should identify in the covering letter of the application that the clinical trial includes the use of an IVD device.
Where no analytical performance study has been conducted at the time of application, a Tabular Summary should be submitted, and the MHRA will assess it and issue a decision letter.
if no analytical performance study has yet been conducted at time of application, a Tabular Summary (MS Word Document, 35.5 KB) should be submitted.
If the MHRA requests further information and no response arrives within 60 calendar days (or an agreed longer period), the clinical trial application will be rejected.
if the MHRA issues a request for further information and no response is received from the applicant within 60 calendar days (or a longer period, if an extension is agreed with MHRA), the clinical trial application will be rejected.
Who should care
The guidance applies to sponsors and applicants for clinical trials with sites in Great Britain, including trials where samples are tested outside the UK but results drive patient management decisions in Great Britain.
This guidance applies to clinical trial sites in Great Britain and includes trials where patient samples are tested outside the United Kingdom, but where the results are used in patient management decisions in Great Britain.
- Clinical trials that include an in vitro diagnostic deviceMHRA, the document itself
- Agency
- UK Medicines and Healthcare products Regulatory Agency
- Type
- Publication
- Last published
- Mar 9, 2026
Status, dates, docket and comment deadline are read from MHRA's own records. The summary is written by a language model and checked by matching each quoted passage against the document; it is a reading aid, not the document.