Clinical investigations: investigators' responsibilities

UK Medicines and Healthcare products Regulatory Agency, publication. In the week of Mar 30 to Apr 5, 2026.

  • Last published: Mar 30, 2026

Summary

Each point with the passage it rests on

What it says

  1. A manufacturer that needs clinical data to support UKCA or CE marking must apply to the MHRA before the clinical investigation is due to begin, and the investigation may only proceed if the MHRA raises no grounds for objection within the statutory review time.

    If such an investigation is necessary, the manufacturer must make an application to the MHRA before the investigation is due to begin.
  2. The manufacturer or sponsor, not the clinical investigator, is responsible for notifying the MHRA and submitting the required documentation, and the clinical investigator will normally have no direct contact with the MHRA.

    It is the responsibility of the manufacturer or the sponsor of the study to both notify the MHRA and to submit the documentation required by the UK MDR 2002 or EU MDR to the MHRA.
  3. Sponsors and investigators must conduct the investigation in full accordance with the approved clinical investigation plan, and any deviations from that plan must be reported to the MHRA.

    Sponsors and investigators must ensure that the investigation is conducted in full accordance with the approved clinical investigation plan ( CIP ) and the requirements of the UK MDR 2002 or EU MDR . Deviations from the approved CIP must be reported to MHRA.
  4. All devices intended for clinical investigation must bear the wording 'exclusively for clinical investigation' and should not bear UKCA, CE or CE marking.

    All devices intended for clinical investigation must bear the wording ‘exclusively for clinical investigation’.
  5. Any proposed changes to the clinical investigation, and any adverse incident involving the device, must be notified or reported to the MHRA.

    All proposed changes to the clinical investigation whether relating to the device, aspects of the clinical investigation plan, investigators or investigating institutions must be notified to the MHRA.

Who should care

  1. Clinical investigators, sponsors and manufacturers of medical devices undergoing clinical investigation in the UK should read this guidance.

    This guidance helps clinical investigators understand what needs to be included in a submission notifying the MHRA of a manufacturer’s plans to undertake a clinical investigation

Sources

From MHRA

Record

Agency
UK Medicines and Healthcare products Regulatory Agency
Type
Publication
Last published
Mar 30, 2026

Status, dates, docket and comment deadline are read from MHRA's own records. The summary is written by a language model and checked by matching each quoted passage against the document; it is a reading aid, not the document.

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.