Clinical investigations in Great Britain

UK Medicines and Healthcare products Regulatory Agency, guidance. In the week of Jul 20 to 26, 2026.

  • Last published: Jul 23, 2026

Summary

Each point with the passage it rests on

What it says

  1. Before devices intended for clinical investigation in Great Britain are made available to a medical practitioner, the manufacturer or UK Responsible Person must give 60 days of prior notice to the Secretary of State for Health by writing to the MHRA.

    Before devices intended for clinical investigation in Great Britain are made available to a medical practitioner for the purposes of clinical investigation, the manufacturer or UK Responsible Person (UKRP) of the device must give 60 days of prior notice to the Secretary of State for Health by writing to the MHRA.
  2. Manufacturers must record fully all adverse events and report all serious adverse events occurring in all participating centres to the MHRA.

    require manufacturers to record fully all adverse events and report all serious adverse events occurring in all participating centres to the MHRA.
  3. Manufacturers must notify the MHRA of all study deviations as soon as they are made aware of them, using the MHRA protocol deviation tracker.

    Manufactures must notify the MHRA of all deviations to the study as soon as they have been made aware of them.
  4. Where a REC opinion is required, a clinical investigation of an unmarked device should not start until both the relevant REC opinion and the MHRA have raised no grounds for objection.

    No clinical investigation of an unmarked device should be started until both the relevant REC opinion and the MHRA have raised no grounds for objection.
  5. Manufacturers must provide the MHRA with a copy of the final clinical investigation report within 1 year of the end of the clinical investigation.

    You should provide us with a copy of the final clinical investigation report within 1 year of the end of the clinical investigation.

Who should care

  1. The guidance is for manufacturers undertaking a clinical investigation for a medical device, and it applies in Great Britain.

    This guidance is for manufacturers undertaking a clinical investigation for a medical device. All of the information applies in Great Britain

Sources

From MHRA

Record

Agency
UK Medicines and Healthcare products Regulatory Agency
Type
Guidance
Last published
Jul 23, 2026

Status, dates, docket and comment deadline are read from MHRA's own records. The summary is written by a language model and checked by matching each quoted passage against the document; it is a reading aid, not the document.