Applying Human Factors and Usability Engineering to Medical Devices: Guidance for Industry and Food and Drug Administration Staff

US Food and Drug Administration CDRH, final guidance. In the week of Aug 3 to 9, 2026.

  • Issued: Aug 3, 2026

Summary

Each point with the passage it rests on

What it says

  1. Human factors validation testing should use test participants who represent the intended users, perform all critical tasks, use the final user interface design, and use test conditions realistic enough to represent actual use.

    The human factors validation testing should be designed as follows:
  2. Design modifications to the user interface are generally the most effective way to eliminate or reduce use-related hazards, and are preferred over labeling or training.

    Design modifications to the device and its user interface are generally the most effective means for eliminating or reducing use-related hazards.
  3. The guidance recommends a minimum of 15 test participants for human factors validation testing, with at least 15 participants for each distinct user population.

    in general, the minimum number of participants should be 15.

Who should care

  1. Medical device manufacturers developing new devices, and FDA staff, are the intended readers of this guidance on applying human factors and usability engineering.

    FDA has developed this guidance document to assist industry in following appropriate human factors and usability engineering processes to maximize the likelihood that new medical devices will be safe and effective for the intended users, uses and use environments.

Sources

From FDA and the Federal Register

Record

Agency
US Food and Drug Administration CDRH
Type
Guidance
Status
Final
Issued
Aug 3, 2026
Text changed
Nov 24, 2026; the summary was rewritten

Status, dates, docket and comment deadline are read from FDA's own records and the docket's notice on Regulations.gov or in the Federal Register. The summary is written by a language model and checked by matching each quoted passage against the document; it is a reading aid, not the document.