Computed Tomography X-ray System
Recall Z-3277-2026, part of FDA recall event 99731. See Shanghai United Imaging Healthcare's company page.
Shanghai United Imaging Healthcare Co., Ltd. started this Class II device recall on Aug 11, 2026 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Aug 11, 2026
- Status: Ongoing
Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
- Product
- Computed Tomography X-ray System, Model: uCT 530 with software
- Quantity
- 4
- Lots and codes
- UDI-DI: 06971576831012. Serial Numbers/Software: 300174/R001.19.5.1839204-Re3300SP1H1_P4, 304006/R001.17.6.1607336-Re3100SP4, 304007/R001.19.4.1810621-Re3300SP1H1_P3, 304008/R001.19.4.1810621-Re3300SP1H1_P3
- Distribution
- US Nationwide distribution in the states of CA, PR, UT, WV, ND, NE, NC, TX, ID, LA, MT, FL, CO, IN, KY, AL, OH, MI, IA, MO, NY, NV, TN, WV, SC, WI, IL, MS, MA, NJ.
- Recalling firm
- Shanghai United Imaging Healthcare Co., Ltd., Shanghai, China
- How the firm notified
- —
- FDA classified
- Sep 25, 2026