Platinium CRT DF1 1711
Recall Z-3230-2017, part of FDA recall event 77983. See Sorin Group Italia CRF's company page.
Sorin Group Italia SRL - CRF started this Class II device recall on Jul 24, 2017 (voluntary: firm initiated); FDA lists it as terminated, terminated Jan 29, 2024.
- Classification: Class II
- Started: Jul 24, 2017
- Status: Terminated
- Products in this event: 6
There is a possibility of overconsumption of certain PLATINIUM Implantable Cardiac Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillators (CRT-Ds) following an ElectroStatic Discharge (ESD) during the implant surgery or a Magnetic Resonance Imaging (MRI) scan, and what actions to take.
- Product
- Platinium CRT DF1 1711, Model Number TDF033U, UDI/GTIN 08031527015538, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single chamber arrhythmia detection Product Usage: PLATINIUM VR is indicated for use in patients who are at high risk of sudden cardiac death due to ventricular tachyarrhythmias and who have experienced one of the following situations: - Survival of at least one episode of cardiac arrest (manifested by the loss of consciousness) due to ventricular tachyarrhythmia, - Recurrent, poorly tolerated sustained ventricular tachycardia (VT).
- Quantity
- 154 units
- Lots and codes
- All Serial Numbers
- Distribution
- US Nationwide Distribution
- Recalling firm
- Sorin Group Italia SRL - CRF, Saluggia, VC, N/A, Italy
- How the firm notified
- Letter
- FDA classified
- Sep 27, 2017
- Platinium VR DF4 1240, Model Number TDF035U, UDI/GTIN 08031527015460, Implantable cardioverter defibrillator, …Z-3231-2017 · Class II · Jul 24, 2017 · Terminated
- Platinium CRT-D 1741, Model Number TDF037U, UDI/GTIN 08031527015484, Implantable cardioverter defibrillator, …Z-3233-2017 · Class II · Jul 24, 2017 · Terminated
- Platinium DR DF4 1540, Model Number TDF036U, UDI/GTIN 08031527015477, Implantable cardioverter defibrillator, …Z-3232-2017 · Class II · Jul 24, 2017 · Terminated
- Platinium VR 1210, Model Number TDF031U, UDI/GTIN 08031527015514, Implantable cardioverter defibrillator, …Z-3228-2017 · Class II · Jul 24, 2017 · Terminated
- Platinium DR 1510, Model Number TDF032U, UDI/GTIN 08031527015521, Implantable cardioverter defibrillator, …Z-3229-2017 · Class II · Jul 24, 2017 · Terminated
- PARADYM Models: VR 8250, DR 8550, CRT-D 8750, RF VR 9250, RF DR 9550, RF CRT-D 9750, SonR 8770, RF SonR 9770 …Z-2113-2017 · Class II · Apr 11, 2017 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports