Arca
Skip to content
Tools
Tools

5008X HD/HDF with Twister Blood Tubing Set Part Number 03-5350-8

Recall Z-3087-2026, part of FDA recall event 99556. See Fresenius Medical Care Renal Therapies Group's company page.

Fresenius Medical Care Renal Therapies Group, LLC started this Class II device recall on Aug 10, 2026 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Aug 10, 2026
  • Status: Ongoing
  • Products in this event: 4

Why It Was Recalled

Class II

Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.

Record

Product
5008X HD/HDF with Twister Blood Tubing Set Part Number 03-5350-8
Quantity
6,624 units
Lots and codes
Lots 25JR01802 25NR01017 25NR01266 UDI-DI 00840861102525
Distribution
Worldwide - US Nationwide distribution in the states of AL, CA, CO, FL, GA, IL, KY, LA, MA, MD, MN, MO, NC, NH, NH, NY, OH, PA, TN, TX, VA and the country of Lithuania.
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC, Waltham, MA, United States
How the firm notified
N/A
FDA classified
Sep 2, 2026

Other Products in This Recall

Recall event 99556

Fresenius Medical Care Renal Therapies Group, LLC's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports