VidiStar(TM) PACS & DICOM Viewer Software system
Recall Z-2992-2018, part of FDA recall event 80896. See Hitachi Healthcare Americas Corp Informatics Division's company page.
Hitachi Healthcare Americas Corp Informatics Division started this Class II device recall on Aug 28, 2018 (voluntary: firm initiated); FDA lists it as terminated, terminated Feb 11, 2021.
- Classification: Class II
- Started: Aug 28, 2018
- Status: Terminated
The secure filesystem client software used in the interface between the Vidistar PACS and an EHR system may cause intermixed images from multiple patients showing in a single study.
- Product
- VidiStar(TM) PACS & DICOM Viewer Software system
- Quantity
- 12 units
- Lots and codes
- versions 3.10.5 and 3.10.6
- Distribution
- SC, CO, IL, AZ, MT, OH, IN, TX, GA, MA
- Recalling firm
- Hitachi Healthcare Americas Corp Informatics Division, Greenville, SC, United States
- How the firm notified
- FDA classified
- Sep 6, 2018
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports