GE Healthcare Optima CT540. The systems are intended for head
Recall Z-2872-2018, part of FDA recall event 80563. See GE Healthcare Japan's company page.
GE Healthcare Japan Corporation started this Class II device recall on Apr 27, 2018 (voluntary: firm initiated); FDA lists it as terminated, terminated Aug 21, 2020.
- Classification: Class II
- Started: Apr 27, 2018
- Status: Terminated
- Products in this event: 4
GE Healthcare has identified that some CT systems may have a damaged cable that can expose 120VAC to Service Engineers working inside the gantry.
- Product
- GE Healthcare Optima CT540. The systems are intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications.Sold under the following product names: BEIJING CT OPTIMA 540, BEIJING OPTIMA, BEIJING OPTIMA 540, BEIJING OPTIMA 540 ASIR, CT BEIJING OPTIMA 540, CT BEIJING OPTIMA 540 ASIR, CT GOLDSEAL OPTIMA 540, CT OPTIMA 540, CT OPTIMA 540 ASIR, CT OPTIMA CT540 TANG BEIJING, DISCOVERY 670 3/8 MEA PRO, HINO OPTIMA 540, HINO OPTIMA 540 ASIR, HINO OPTIMA 540 ASIR, OPTIMA 540, OPTIMA 540 ASIR, OPTIMA 540 MOBILE BEIJING, OPTIMA CT540, OPTIMA CT540 32 SLICES, OPTIMA CT540 ASIR, OPTIMA CT540 BJG SKD RUSSIA, Optima CT540 Gantry Top Asm with Rating Plate,SGB tube and CJ cover, Optima CT540 Gantry Top Asm with Rating Plate,SGB tube and CJ cover for Hino, OPTIMA CT540 TANG, OPTIMA CT540 TANG BEIJING, OPTIMA CT540 TANG HINO
- Quantity
- 4,007 devices total (products 1, 2, 3)
- Lots and codes
- Model Number, Serial Number (System ID): 5432539, 00000409871HM1 (IR1304CT02); 5432539, 00000406425HM9 (IR1563CT01); 5432539-2, Not Available (970GMCT1); 5432539-2, 343769HM6 (262434OPTCT1); 5432539, 383019HM7 (251809DWMCT); 5432539-2, 00000058405YC2 (928289CT); 5432539, 00000340698HM0 (TMCACT540); 5432539, 340698HM0 (AM16OG71165); 5432539, 377682HM0 (850314CT); 5432539, 00000333256HM6 (954916OPT16); 5432539, 00000350436HM2 (912466OCT); 5432539, 00000354310HM5 (770382ERCT); 5432539, 00000361670HM3 (919956CT); 5432539, 00000370798HM1 (704403KAN); 5432539, 00000358402HM6 (910642CT2); 5432539,…
- Distribution
- Worldwide Distribution: US (Nationwide) distribution to states of: AK, AL, CA, CO, CT, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, RI, SC, SD, TN, TX, UT, WI, WV, WY; and countries of:: ALGERIA, ARGENTINA, ARMENIA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BOLIVIA, BOSNIA AND HERZEGOVINA, BRAZIL, BULGARIA, CAMEROON, Canada, CHILE, CHINA, COLOMBIA, Cote D'Ivoire, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, ETHIOPIA, FINLAND, FRANCE, French Guiana, GERMANY, GHANA, GREECE, GUADELOUPE, GUATEMALA, HONDURAS, Hong Kong, HUNGARY, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, Korea (Republic of), KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIA, MALAYSIA, Mali, MALTA, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NEW , ZEALAND, NIGERIA, NORWAY, OMAN, PAKISTAN, Palestine, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL, PUERTO RICO, REUNION, ROMANIA, Russia, RWANDA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TANZANIA, THAILAND, TUNISIA, TURKEY, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, UZBEKISTAN, VENEZUELA, VIETNAM, and YEMEN.
- Recalling firm
- GE Healthcare Japan Corporation, Hino, Japan
- How the firm notified
- Letter
- FDA classified
- Aug 17, 2018
- GE Healthcare Optima CT660. The systems are intended for head, whole body, cardiac and vascular X-ray Computed…Z-2873-2018 · Class II · Apr 27, 2018 · Terminated
- GE Healthcare Optima CT660. Sold under the following product names: BEIJING OPTIMA CT660 PRO GT170, BEIJING…Z-2871-2018 · Class II · Apr 27, 2018 · Terminated
- GE Healthcare Revolution EVO, Optima CT660, Optima CT670, Optima CT680. The systems are intended for head, whole body,…Z-2870-2018 · Class II · Apr 27, 2018 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports