TriMed Semi-Tubular Plate
Recall Z-2849-2016, part of FDA recall event 75080. See TriMed's company page.
Trimed Inc started this Class II device recall on Aug 29, 2016 (voluntary: firm initiated); FDA lists it as terminated, terminated Nov 1, 2016.
- Classification: Class II
- Started: Aug 29, 2016
- Status: Terminated
TriMed is recalling the Semi-Tubular Plates with 10 holes (SMTP-10) because a non-locking screw passed through a hole in a SMTP-10 plate being implanted.
- Product
- TriMed Semi-Tubular Plate, 10-Hole, Model No. SMTP-10 Contoured plate with offset screw holes. The plate is 103 mm in length, 1.3 mm thick and 9.6 mm wide. The part is made of surgical steel conformance with ASTM F138 or F139. The plate is used to fixation of long bone fractures. The TriMed Bone Plates are intended for use in the fixation of fractures to the Tibia, Fibula, Ulna, Radius and the Humerus.
- Quantity
- 30 units
- Lots and codes
- Lot no. 06206
- Distribution
- Worldwide Distribution-US (nationwide) including states of: Alabama, Florida, Indiana, Connecticut, California, Wisconsin, Missouri, Florida, Hawaii; and countries of: Sweden and South Africa.
- Recalling firm
- Trimed Inc, Santa Clarita, CA, United States
- How the firm notified
- Letter
- FDA classified
- Sep 20, 2016