NobelReplace Conical Connection (CC) PMC RP 4.3x10mm
Recall Z-2828-2020, part of FDA recall event 86085. See Nobel Biocare Services's company page.
NOBEL BIOCARE SERVICES AG started this Class II device recall on Jul 20, 2020 (voluntary: firm initiated); FDA lists it as terminated, terminated May 12, 2021.
- Classification: Class II
- Started: Jul 20, 2020
- Status: Terminated
Dental implant cap label may incorrectly indicate implant measurements 5.0x8mm. Actual implant measurements are 4.3x10mm. If not noticed prior to implant placement, this could lead to: 1) A second surgical procedure; 2) Implant replacement; 3) If placed in posterior lower jaw, the implant may damage the inferior alveolar nerve canal or perforate a small artery requiring immediate implant removal.
- Product
- NobelReplace Conical Connection (CC) PMC RP 4.3x10mm, REF: 37291 - Product Usage: implants are intended to be used in the upper or lower jaw bone (osseointegration) and used for anchoring or supporting tooth replacements to restore chewing function.
- Quantity
- 188
- Lots and codes
- Lot: 13084072, UDI: (01)07332747073584(10)13084072(17)240525
- Distribution
- Worldwide distribution - U.S. Nationwide distribution including in the states of FL and NY and the countries of Albania and Hong Kong.
- Recalling firm
- NOBEL BIOCARE SERVICES AG, Kloten, N/A, Switzerland
- How the firm notified
- Letter
- FDA classified
- Aug 20, 2020