Arca
Skip to content
Tools
Tools

NobelReplace Conical Connection (CC) PMC RP 4.3x10mm

Recall Z-2828-2020, part of FDA recall event 86085. See Nobel Biocare Services's company page.

NOBEL BIOCARE SERVICES AG started this Class II device recall on Jul 20, 2020 (voluntary: firm initiated); FDA lists it as terminated, terminated May 12, 2021.

  • Classification: Class II
  • Started: Jul 20, 2020
  • Status: Terminated

Why It Was Recalled

Class II

Dental implant cap label may incorrectly indicate implant measurements 5.0x8mm. Actual implant measurements are 4.3x10mm. If not noticed prior to implant placement, this could lead to: 1) A second surgical procedure; 2) Implant replacement; 3) If placed in posterior lower jaw, the implant may damage the inferior alveolar nerve canal or perforate a small artery requiring immediate implant removal.

Record

Product
NobelReplace Conical Connection (CC) PMC RP 4.3x10mm, REF: 37291 - Product Usage: implants are intended to be used in the upper or lower jaw bone (osseointegration) and used for anchoring or supporting tooth replacements to restore chewing function.
Quantity
188
Lots and codes
Lot: 13084072, UDI: (01)07332747073584(10)13084072(17)240525
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of FL and NY and the countries of Albania and Hong Kong.
Recalling firm
NOBEL BIOCARE SERVICES AG, Kloten, N/A, Switzerland
How the firm notified
Letter
FDA classified
Aug 20, 2020

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports