Arca
Skip to content
Tools
Tools

Axiom 2.8 implants Sterile The Axiom REG implants are used as artificial root structures…

Recall Z-2754-2016, part of FDA recall event 74722. See Anthogyr's company page.

Anthogyr started this Class II device recall on Jun 20, 2016 (voluntary: firm initiated); FDA lists it as terminated, terminated Mar 28, 2017.

  • Classification: Class II
  • Started: Jun 20, 2016
  • Status: Terminated
  • Products in this event: 2

Why It Was Recalled

Class II

Following a FDA inspection showing that the firm failed to validate the design and process of its implants.

Record

Product
Axiom 2.8 implants Sterile The Axiom REG implants are used as artificial root structures for replacement of missing teeth. They can be used for stabilization of removable prostheses or fixation of single tooth restorations or partial dentures. There are different diameters and different lengths for each diameter.
Quantity
10 units
Lots and codes
Code OP28100 Size 02.8 x 10.0 Code OP28120 Size 02.8 x 12.0 Code OP28140 Size 02.8 x 14.0
Distribution
US Distribution to sates of: NC, NY, and OR.
Recalling firm
Anthogyr, Sallanches, France
How the firm notified
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
FDA classified
Sep 8, 2016

Other Products in This Recall

Recall event 74722