Temporary Total Artificial Heart (TAH-t) with Freedom Driver System REF 595000-001…
Recall Z-2731-2015, part of FDA recall event 71905. See SynCardia Systems's company page.
SynCardia Systems Inc. started this Class I device recall on Aug 6, 2015 (voluntary: firm initiated); FDA lists it as terminated, terminated Nov 4, 2015.
- Classification: Class I
- Started: Aug 6, 2015
- Status: Terminated
SynCardia Systems, Inc. is recalling Freedom Drivers due to a specific component of the drive mechanism may fail and cause the drive mechanism to stop pumping; resulting in the loss of life-sustaining function.
- Product
- Temporary Total Artificial Heart (TAH-t) with Freedom Driver System REF 595000-001 Product Usage: Freedom Drivers are components of the SynCardia temporary Total Artificial Heart (TAH-t) System, which is indicated for use in transplant-eligible candidates at risk of imminent death from biventricular failure. Freedom Drivers are intended for use in and out of the hospital.
- Quantity
- 56 drivers (21 currently in use on patients)
- Lots and codes
- Part Number: 595000-001 Motor Gearbox Serial # CGI Lot # SynCardia Lot # Freedom Driver Serial # 85979-013 85979 031479 4506 85979-022 85979 031479 4506 85977-033 85977 031333 4507 85976-020 85976 031041 4514 85976-010 85976 030944 4517 85976-014 85976 030944 4517 85979-014 85979 031479 4517 85979-021 85979 031479 4517 85976-039 85976 031074 4521 85977-001 85977 031074 4521 85978-034 85978 031823 4521 85978-040 85978 031823 4521 85977-020 85977 031333 4542 85977-032 85977 031333 4542 85979-005 85979 031452 4546 85979-012 85979 031452 4546 85979-018 85979 031479 4559 85979-024 85979 031479…
- Distribution
- Worldwide Distribution -- US Nationwide in the states of AR, NC, CA, OH, PA, TX, MI, and VA; and the countries of Germany, Turkey and France.
- Recalling firm
- SynCardia Systems Inc., Tucson, AZ, United States
- How the firm notified
- Letter
- FDA classified
- Sep 17, 2015
Newest first, one product per recall event; search them all