GE Healthcare AW Server 3.2 ext.6.5; Radiological Image Processing System
Recall Z-2553-2026, part of FDA recall event 99042. See GE Medical Systems SCS's company page.
GE Medical Systems SCS started this Class II device recall on May 8, 2026 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: May 8, 2026
- Status: Ongoing
GE HealthCare has become aware of a context synchronization issue in AW Server 3.2 ext. 6.5. When a user selects a patient or exam in the AW Server Web Client worklist and launches an interactive application (e.g. Volume Viewer), the application may open the previous patient's exam instead of the intended one. When this issue occurs, there is no system warning or error notification. If the error is not recognized, a clinical user could review, interpret, or report images for the wrong patient which could lead to misdiagnosis, incorrect clinical decisions, resulting in delayed or incorrect treatment.
- Product
- GE Healthcare AW Server 3.2 ext.6.5; Radiological Image Processing System
- Quantity
- 340 servers
- Lots and codes
- REF# (01)00840682102384(10)AWS03D02E6D5, GTIN Number 00840682102384
- Distribution
- Worldwide distribution - United States Nationwide and the countries of Algeria, Australia, Belarus, Belgium, Benin, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Denmark, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Korea (Republic Of), Lithuania, Morocco, Nepal, Netherlands, New Zealand, North Macedonia, Norway, Pakistan, Poland, Portugal, Romania, Russia, Saudi Arabia, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkiye, United Arab Emirates, United Kingdom, Vietnam.
- Recalling firm
- GE Medical Systems SCS, Buc, N/A, France
- How the firm notified
- Letter
- FDA classified
- Jun 24, 2026
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports