Arca
Skip to content
Tools
Tools

GE Healthcare AW Server 3.2 ext.6.5; Radiological Image Processing System

Recall Z-2553-2026, part of FDA recall event 99042. See GE Medical Systems SCS's company page.

GE Medical Systems SCS started this Class II device recall on May 8, 2026 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: May 8, 2026
  • Status: Ongoing

Why It Was Recalled

Class II

GE HealthCare has become aware of a context synchronization issue in AW Server 3.2 ext. 6.5. When a user selects a patient or exam in the AW Server Web Client worklist and launches an interactive application (e.g. Volume Viewer), the application may open the previous patient's exam instead of the intended one. When this issue occurs, there is no system warning or error notification. If the error is not recognized, a clinical user could review, interpret, or report images for the wrong patient which could lead to misdiagnosis, incorrect clinical decisions, resulting in delayed or incorrect treatment.

Record

Product
GE Healthcare AW Server 3.2 ext.6.5; Radiological Image Processing System
Quantity
340 servers
Lots and codes
REF# (01)00840682102384(10)AWS03D02E6D5, GTIN Number 00840682102384
Distribution
Worldwide distribution - United States Nationwide and the countries of Algeria, Australia, Belarus, Belgium, Benin, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Denmark, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Korea (Republic Of), Lithuania, Morocco, Nepal, Netherlands, New Zealand, North Macedonia, Norway, Pakistan, Poland, Portugal, Romania, Russia, Saudi Arabia, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkiye, United Arab Emirates, United Kingdom, Vietnam.
Recalling firm
GE Medical Systems SCS, Buc, N/A, France
How the firm notified
Letter
FDA classified
Jun 24, 2026

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports