Patient Positioning System with KRC2 controller using software versions 2.3.1 2.3.10…
Recall Z-2537-2018, part of FDA recall event 80510. See Forte Automation Systems's company page.
Forte Automation Systems Inc started this Class II device recall on Sep 5, 2017 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 13, 2020.
- Classification: Class II
- Started: Sep 5, 2017
- Status: Terminated
Communications between the Patient Positioning System and the accuracy filter can periodically fail with no clear indication to the operator.
- Product
- Patient Positioning System with KRC2 controller using software versions 2.3.1 2.3.10 Product Usage: The patient positioning system is a SCARA designed robotic arm designed to position a patient for medical procedures prescribed by oncologists and others that require a high degree of accuracy and repeatability.
- Quantity
- 13
- Lots and codes
- PPS Serial Number Cabinet Type Software Version 7757-11 KRC2 2.3.3 7757-21 KRC2 2.3.3 7757-31 KRC2 2.3.3 7757-41 KRC2 2.3.3 7757-51 KRC2 2.3.8 7755-21 KRC2 2.3.3 7755-31 KRC2 2.3.3 7755-41 KRC2 2.3.3 7392-2 KRC2 2.3.3 7392-3 KRC2 2.3.3 7392-4 KRC2 2.3.3 7392-1 KRC2 2.3.8 7392-5 KRC2 2.3.8
- Distribution
- US in the state of California
- Recalling firm
- Forte Automation Systems Inc, Machesney Park, IL, United States
- How the firm notified
- Letter
- FDA classified
- Jul 24, 2018