Patient Positioning System with KRC2 controller using software versions 2.3.1 2.3.10…

Recall Z-2537-2018, part of FDA recall event 80510. See Forte Automation Systems's company page.

Forte Automation Systems Inc started this Class II device recall on Sep 5, 2017 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 13, 2020.

  • Classification: Class II
  • Started: Sep 5, 2017
  • Status: Terminated

Why It Was Recalled

Class II

Communications between the Patient Positioning System and the accuracy filter can periodically fail with no clear indication to the operator.

Record

Product
Patient Positioning System with KRC2 controller using software versions 2.3.1 2.3.10 Product Usage: The patient positioning system is a SCARA designed robotic arm designed to position a patient for medical procedures prescribed by oncologists and others that require a high degree of accuracy and repeatability.
Quantity
13
Lots and codes
PPS Serial Number Cabinet Type Software Version 7757-11 KRC2 2.3.3 7757-21 KRC2 2.3.3 7757-31 KRC2 2.3.3 7757-41 KRC2 2.3.3 7757-51 KRC2 2.3.8 7755-21 KRC2 2.3.3 7755-31 KRC2 2.3.3 7755-41 KRC2 2.3.3 7392-2 KRC2 2.3.3 7392-3 KRC2 2.3.3 7392-4 KRC2 2.3.3 7392-1 KRC2 2.3.8 7392-5 KRC2 2.3.8
Distribution
US in the state of California
Recalling firm
Forte Automation Systems Inc, Machesney Park, IL, United States
How the firm notified
Letter
FDA classified
Jul 24, 2018

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.