TruDi NAV Suction, 0 Degrees; Model/Catalog Number: TDNS000Z. TruDi" NAV Suction…
Recall Z-2507-2024, part of FDA recall event 94955. See Integra LifeSciences's company page.
Integra LifeSciences Corp. started this Class II device recall on Jul 9, 2024 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Jul 9, 2024
- Status: Ongoing
Specific lots of TruDi NAV Suction Instruments 0 Degrees were incorrectly calibrated which when used with the TruDi Navigation System, may result in a discrepancy between the actual tip location and the visualized location displayed on the system.
- Product
- TruDi NAV Suction, 0 Degrees; Model/Catalog Number: TDNS000Z. TruDi" NAV Suction Instruments are intended for use with the TruDi" Navigation System during surgical procedures in ENT medicine and skull base surgery
- Quantity
- 265 units
- Lots and codes
- UDI/DI: 10705031245877; Lot Codes: 2310160, 2309110, 2309150, 2308140, 2308160, 2308180.
- Distribution
- The devices are only sold in the US. The following states have received the product: AL, AZ CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, PA, SC, SD, TN, TX, VA, VT, WI.
- Recalling firm
- Integra LifeSciences Corp., Princeton, NJ, United States
- How the firm notified
- Letter
- FDA classified
- Aug 5, 2024
Newest first, one product per recall event; search them all
- AURORA Surgiscope System, Sterile, single use device that contains a Sheath, Obturator, and Imager. Manufacturer's…Z-1295-2025 · Class II · Feb 6, 2025 · Ongoing
- Brand Name: MediHoney¿ Gel Product Name: MediHoney¿ Wound & Burn Dressing Gel Model/Catalog Number: 31815 Software…Z-1168-2025 · Class II · Jan 10, 2025 · Ongoing
- CODMAN CERTAS Plus Programmable Valve Inline Small Valve Only. Model Number: 828810PL. Implantable device that…Z-0978-2025 · Class II · Dec 16, 2024 · Ongoing
- Duraform Dural Graft Implant, Size 3x3 , Model Number 80-1477US Lot CT007694Z-0262-2025 · Class II · Sep 26, 2024 · Ongoing
- SURG PAT XRAY 3/4X3/4 Model/Catalog Number: 801401. The surgical patties and surgical strips are indicated for the…Z-0240-2025 · Class II · Sep 20, 2024 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports