Arca
Skip to content
Tools
Tools

GS70 Salus Surgical Table

Recall Z-2469-2023, part of FDA recall event 92290. See Skytron's company page.

Skytron, LLC started this Class II device recall on Aug 1, 2023 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Aug 1, 2023
  • Status: Ongoing
  • Products in this event: 2

Why It Was Recalled

Class II

When the control of the operating table was used in Bluetooth mode, the operating table continued moving even after releasing the button on the control. Originally, the root cause was thought to be confined to Bluetooth/wireless communications between the control pendant and the surgical table. However, during investigation, Mizuho discovered that the failure could be replicated even in wired mode if communication between the pendant and table were somehow interrupted. The root cause appears to be: if communication between the controller and table is interrupted while a button is pressed, the table does not know if/when a button is released. This occurs in wireless or wired mode.

Record

Product
GS70 Salus Surgical Table, Product Code 18-070-70
Quantity
110 units
Lots and codes
Serial Numbers: SUC-20207-0006 SUC-20208-0025 SUC-20209-0044 SUC-20212-0063 SUC-20228-0082 SUC-2022Y-0101 SUC-20207-0007 SUC-20208-0026 SUC-20209-0045 SUC-20212-0064 SUC-20228-0083 SUC-2022Y-0102 SUC-20207-0008 SUC-20208-0027 SUC-2020X-0046 SUC-20212-0065 SUC-20228-0084 SUC-2022Y-0103 SUC-20207-0009 SUC-20208-0028 SUC-2020X-0047 SUC-20226-0066 SUC-20228-0085 SUC-2022Y-0104 SUC-20207-0010 SUC-20208-0029 SUC-2020X-0048 SUC-20226-0067 SUC-20228-0086 SUC-2022Y-0105 SUC-20207-0011 SUC-20209-0030 SUC-2020X-0049 SUC-20226-0068 SUC-20228-0087 SUC-20231-0106 SUC-20207-0012 SUC-20209-0031…
Distribution
Worldwide - US Nationwide distribution in the states of AL, AR, CA, CT, FL, GA, IA, IL, IN, MA, MI, MO, PA, TN, TX, UT, WA, WI and the countries of Australia, New Zealand.
Recalling firm
Skytron, LLC, Grand Rapids, MI, United States
How the firm notified
Letter
FDA classified
Aug 28, 2023

Other Products in This Recall

Recall event 92290