Integra Jarit GEMINI CLAMP 9 1/4 inch - used in vascular and cardiothoracic procedures…
Recall Z-2435-2019, part of FDA recall event 83418. See Integra LifeSciences's company page.
Integra LifeSciences Corp. started this Class II device recall on Aug 7, 2019 (voluntary: firm initiated); FDA lists it as terminated, terminated Jun 24, 2020.
- Classification: Class II
- Started: Aug 7, 2019
- Status: Terminated
A small crack at the box lock corner may after multiple cleaning and sterilization cycles produce an increased risk of corrosion, weakening the instrument and resulting in premature need for replacement and/or breakage of the instrument
- Product
- Integra Jarit GEMINI CLAMP 9 1/4 inch - used in vascular and cardiothoracic procedures used for clamping and occluding vessels. Cat. #140-332
- Quantity
- 237 units
- Lots and codes
- Lot #AA1901
- Distribution
- US Nationwide distribution.
- Recalling firm
- Integra LifeSciences Corp., Plainsboro, NJ, United States
- How the firm notified
- Letter
- FDA classified
- Aug 30, 2019
Newest first, one product per recall event; search them all
- AURORA Surgiscope System, Sterile, single use device that contains a Sheath, Obturator, and Imager. Manufacturer's…Z-1295-2025 · Class II · Feb 6, 2025 · Ongoing
- Brand Name: MediHoney¿ Gel Product Name: MediHoney¿ Wound & Burn Dressing Gel Model/Catalog Number: 31815 Software…Z-1168-2025 · Class II · Jan 10, 2025 · Ongoing
- CODMAN CERTAS Plus Programmable Valve Inline Small Valve Only. Model Number: 828810PL. Implantable device that…Z-0978-2025 · Class II · Dec 16, 2024 · Ongoing
- Duraform Dural Graft Implant, Size 3x3 , Model Number 80-1477US Lot CT007694Z-0262-2025 · Class II · Sep 26, 2024 · Ongoing
- SURG PAT XRAY 3/4X3/4 Model/Catalog Number: 801401. The surgical patties and surgical strips are indicated for the…Z-0240-2025 · Class II · Sep 20, 2024 · Ongoing