Sterile Distraction Screws of 3 different sizes: 12mm
Recall Z-2385-2012, part of FDA recall event 62899. See TeDan Surgical Innovations's company page.
Tedan Surgical Innovations Llc started this Class II device recall on Jun 11, 2012 (voluntary: firm initiated); FDA lists it as terminated, terminated Jul 31, 2013.
- Classification: Class II
- Started: Jun 11, 2012
- Status: Terminated
This letter is to notify you that several of our sterile screw products are being voluntarily recalled. Product sterility may be compromised due to a mechanical failure of the protective packaging tube. Tube end plugs may dislodge creating the potential for the screw to come out of the protective tube and then the sharp end of the screw pierce the outer sterile barrier pouch.
- Product
- Sterile Distraction Screws of 3 different sizes: 12mm, 14mm and 16mm sold under TSI, 3COR, DB Surgical, and American Surgical Products brands and Sterile XL Screws of 7 different sizes: 90mm, 100mm, 110mm, 120mm, 130mm, 140mm, 150mm sold under TSI brand Sterile Distraction Screws are intended for the distraction in the anterior approach to the cervical spine. Sterile XL Screws are intended for the fixation of the blade to the vertebral body to prohibit drifting in the extreme lateral approach to the lumbar spine.
- Quantity
- 12,065 Sterile Distraction Screws (2413 boxes); 384 Sterile XL Screws (128 boxes) in distribution
- Lots and codes
- Lot numbers: 019715, 019716, 019942, 019943, 019977, 019978, 020102, 020103, 020104, 020105, 020106, 020107, 020108, 020609, 020782, 021539, 021663, 022097, 022098, 022099, 022122, 022123, 022124, 022371, 022372, 022790, 022791, 022792, 022794, 022795, 022796, 022827, 022828, and 023266.
- Distribution
- Worldwide distribution: USA (nationwide) and countries of : Australia and Taiwan.
- Recalling firm
- Tedan Surgical Innovations Llc, Sugar Land, TX, United States
- How the firm notified
- Letter
- FDA classified
- Sep 14, 2012
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports