ImagePilot, Model No. D9MA
Recall Z-2282-2018, part of FDA recall event 80220. See Konica Minolta Medical Imaging USA's company page.
Konica Minolta Medical Imaging USA, Inc. started this Class II device recall on May 9, 2018 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 10, 2020.
- Classification: Class II
- Started: May 9, 2018
- Status: Terminated
- Products in this event: 2
Panasonic has issued a Mandatory Product Safety Notice, indicating a possibility of battery ignition (fire) in CF-C2 "Toughbook" laptops. Since Konica Minolta uses this laptop in some of its product configurations, it notified its customers to utilize a BIOS Utility for the laptops to minimize the risk of an incident while Panasonic develops a final solution.
- Product
- ImagePilot, Model No. D9MA
- Quantity
- —
- Lots and codes
- All version numbers.
- Distribution
- Nationwide distribution. Foreign distribution to Argentina, Brazil, Canada, Chile, Colombia, Ecuador, Japan, Mexico, Netherlands, and Peru.
- Recalling firm
- Konica Minolta Medical Imaging USA, Inc., Wayne, NJ, United States
- How the firm notified
- Letter
- FDA classified
- Jun 22, 2018
- CS-7, Model No. A4C1Z-2281-2018 · Class II · May 9, 2018 · Terminated
- Sonimage HS1 Ultrasound Kit AC adapter, Product Number: A7AP, UDI: 04560141940031Z-1212-2018 · Class II · Dec 4, 2017 · Terminated