WOM Tube Set for lrrlgatlon
Recall Z-2243-2024, part of FDA recall event 94749. See WOM World of Medicine's company page.
WOM World Of Medicine AG started this Class II device recall on May 24, 2024 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: May 24, 2024
- Status: Ongoing
Incorrect transponder data was written on the RFID tags triggers the error message E03, Tube Set lnvalid, may result in a delay of surgery
- Product
- WOM Tube Set for lrrlgatlon, single-use- Intended to transport irrigation fluid to the patient in endoscopic surgeries. Distr. by Endoscopy Smith & Nephew Inc. REF Numbers: T0449-01
- Quantity
- 1700 units
- Lots and codes
- UDI: 04056702003678 Lot Numbers: 4030902, 4030903
- Distribution
- US Nationwide distribution.
- Recalling firm
- WOM World Of Medicine AG, Ludwigsstadt, N/A, Germany
- How the firm notified
- FDA classified
- Jul 1, 2024
Newest first, one product per recall event; search them all
- Aquilex Fluid Control System-Intended to provide fluid distension of the uterus during diagnostic and operative…Z-2629-2023 · Class II · Jul 17, 2023 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports