The D*Clot¿ HD Rotational Thrombectomy System is a percutaneous catheter-based system…
Recall Z-2213-2024, part of FDA recall event 94741. See Mermaid Medical A/S's company page.
Mermaid Medical A/S started this Class II device recall on May 1, 2024 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: May 1, 2024
- Status: Ongoing
Internal product testing failed bubble leak testing (ASTM F2096), which indicates the sterility barrier may not be fully intact.
- Product
- The D*Clot¿ HD Rotational Thrombectomy System is a percutaneous catheter-based system available in 6 French. The disposable system consists of an outer sheath through which a rotatable radiopaque spiral shaft extends ending in an eccentric blunt distal tip. Once activated, the shaft rotates at approximately 10.000 RPMs to break thrombus into smaller particles, which can be simultaneously aspirated using a dedicated vacuum syringe. The catheter is a stand-alone system and is powered by a 4.5 VDC alkaline battery pack in the control unit.
- Quantity
- 190 devices
- Lots and codes
- Unique Device Identifier: 15711055004333; Lot codes: S2306001; Model Number: T10106065
- Distribution
- Worldwide - US Nationwide distribution in the states of OH, MO, CO, NC, TX, MA, GA, FL, IL, IN, MD, PA, VA, and the country of Chile.
- Recalling firm
- Mermaid Medical A/S, Stenlose, Denmark
- How the firm notified
- FDA classified
- Jun 26, 2024
Newest first, one product per recall event; search them all
- Mermaid Medical M-Biopsy Coaxial Introducer Needle, Coaxial Needle for M-Biopsy 18Gx10cm-PRO-MAG, RX Only. Single…Z-2190-2015 · Class II · Jun 15, 2015 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports