Contour¿ next GEN Blood Glucose Monitoring System
Recall Z-2211-2023, part of FDA recall event 92536. See Ascensia Diabetes Care US's company page.
Ascensia Diabetes Care US, Inc. started this Class II device recall on May 19, 2023 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: May 19, 2023
- Status: Ongoing
Ascensia Diabetes Care has determined through customer complaints that 580 of 2900 meters from lot number DM01T033P are reconfigured from mg/dL to mmol/L. Among them, 579 meters distributed in the United States were packaged into the meter kit and had incorrect factory-set unit of measure where the meters display glucose results in mmol/L rather than mg/dL. The standard unit of measurement for the United States is mg/dL. If a consumer does not notice the incorrect unit of measurement, it is possible that the meters glucose measurement will be read as a lower blood glucose measurement than expected, and this may result in the patients glucose level remaining high, which can potentially lead to an injury requiring immediate medical intervention.
- Product
- Contour¿ next GEN Blood Glucose Monitoring System
- Quantity
- 2,888 units
- Lots and codes
- Model No.: 7917; Lot No. DM01T033P; Serial No. (UDI): 1057018 ((01)00301937917011(17)270901(10)DM01T033P(21)7917-1057018), 1057019 ((01)00301937917011(17)270901(10)DM01T033P(21)7917-1057019), 1057020 ((01)00301937917011(17)270901(10)DM01T033P(21)7917-1057020), 1057023 ((01)00301937917011(17)270901(10)DM01T033P(21)7917-1057023), 1057024 ((01)00301937917011(17)270901(10)DM01T033P(21)7917-1057024), 1057025 ((01)00301937917011(17)270901(10)DM01T033P(21)7917-1057025), 1057076 ((01)00301937917011(17)270901(10)DM01T033P(21)7917-1057076), 1057077…
- Distribution
- Domestic: CA, CO, FL, IN, MO, NJ, NY, OH, RI, SC, TN, TX, & VA.
- Recalling firm
- Ascensia Diabetes Care US, Inc., Valhalla, NY, United States
- How the firm notified
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- FDA classified
- Jul 20, 2023