Recall Z-2195-2013, part of FDA recall event 66110. See Sybron Dental Specialties's company page.

Sybron Dental Specialties started this Class II device recall on Jun 6, 2013 (voluntary: firm initiated); FDA lists it as terminated, terminated Sep 13, 2013.

  • Classification: Class II
  • Started: Jun 6, 2013
  • Status: Terminated

Why It Was Recalled

Class II

The firm initiated the recall of AOA Mini RPE Screw Assembly, because it may be missing a complete weld joining one side of the arm with the screw body. Herbst, MARA and Fixed Metal devices may be affected, because they contain an AOA Mini RPE Screw Assembly.

Record

Product
AOA Mini RPE Screw Assembly , a component of Herbst, MARA and Fixed Metal devices. Product Usage: AOA Mini RPE Screw Assembly is a component of Herbst, MARA and Fixed Metal devices; removable functional appliances that are to modify the growth of the jaws in a prescribed growth pattern, however, this category is generally attached to the patients upper and lower molars or bicuspids by means of temporary stainless steel crowns or orthodontic bands. Retainer, Screw Expansion, Orthodontic.
Quantity
1,091
Lots and codes
The AOA Mini RPE Screw Assembly is a component used in customized palate expander appliances assembled by prescription. The affected assembly may be found in products listed in the table below: Product Description: Herbst Bite Jumper Herbst Appliance Acrylic Splint M.A.R.A. (Mandibular Anterior Reposi tioning Appliance) Rapid Palatal Expander Lower Expander DeLuke Bonded Frameless Expander Haas Expander AOA Mini Rapid Palatal Expander Dischinger Lower Expansion Appliance Tracy/Hilgers MDA Appliance Distal Jet Molar Uprighter Part Number: 3000-0010 3000-0011 3000-0012 3000-0040 3000-0042…
Distribution
Worldwide Distribution - USA Nationwide and the countries of Australia, Canada, and EU.
Recalling firm
Sybron Dental Specialties, Orange, CA, United States
How the firm notified
Letter
FDA classified
Sep 12, 2013

Sybron Dental Specialties's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.